U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA PRIMUS; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA PRIMUS; ANESTHESIA UNITS Back to Search Results
Catalog Number 8603800
Device Problems Gas Output Problem (1266); Failure to Deliver (2338); Intermittent Communication Failure (4038)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/24/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.
 
Event Description
It was reported that the device had a ventilator failure during use.There was no patient injury reported.
 
Event Description
Please refer to the initial report.
 
Manufacturer Narrative
The investigation was performed based on the electronic log file.A malfunction of a cpu board which controls the device-internal communication between user interface and vgc (ventilation and gas controller) was identified to be root cause of the reported failure.This leads to a shutdown of the ventilator and gas mixer simultaneously.In this situation, the device automatically switches to monitoring mode while alarming the user to this condition by means of a corresponding alarm.Manual ventilation with emergency oxygen dosage remains possible including the application of anesthetic gas as well.The procedure how to establish the emergency gas supply is described in the ifu.Dräger finally concludes, that the device has reacted according to its safety concept and has performed an emergency shutdown of the affected components accompanied by the respective alarms.Similar cases are known ¿ however, the exact failure mechanism could not be determined during in-depth analysis.It was only possible to narrow down the root cause to the respective pcb.Thus, it was recommended to replace the affected board.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMUS
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM  23542
MDR Report Key10137136
MDR Text Key194613540
Report Number9611500-2020-00201
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K042607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 08/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8603800
Was Device Available for Evaluation? Yes
Date Manufacturer Received05/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-