Model Number 1458Q/86 |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/28/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This product is registered as a combination product.The results, method, and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
During the implant procedure, the guidewire could not be inserted into the left ventricular (lv) lead.The lv lead was explanted and replaced to resolve the event.The patient was in stable condition.
|
|
Manufacturer Narrative
|
As received, a complete lead was returned in one piece.The guidewire was inserted through the proximal end of the lead all the way through with no restrictions noted.Visual inspection of the lead did not find any anomalies with the exception of procedural damage.The reported event of guidewire could not be inserted was not confirmed.
|
|
Search Alerts/Recalls
|