• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC THREE-WAY PLASTIC STOPCOCK; KGZ ACCESSORIES, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC THREE-WAY PLASTIC STOPCOCK; KGZ ACCESSORIES, CATHETER Back to Search Results
Catalog Number PTWS-2FLL-MLL-R
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/02/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event Description
As reported, prior to a transcatheter arterial chemo embolization procedure, a drug leaked from two three-way plastic stopcocks during the mixing process.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Event Description
No new patient or event information to report.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.As reported, prior to a transcatheter arterial chemo embolization procedure, a drug leaked from two three-way plastic stopcocks during the mixing process.Investigation ¿ evaluation a document based investigation was performed including a review of complaint history, device history record, documentation, drawings, manufacturing instructions, quality control data, and specifications.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was made out of specification.A review of the device history record found no non-conformances related to the reported failure mode.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints that have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.A review of complaint history records shows no other complaints associated with the complaint device lot.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually/functionally inspected by quality control and no related gaps in production or processing controls were noted.Cook has concluded that the conclusion from previously completed root cause investigation aligns with the conclusion of this complaint.The stopcock stems molded of nylon 6 have the potential to absorb moisture resulting in an increase in diameter raising the potential for cracking of the stopcock body contributed to this incident.Per the quality engineering risk assessment, no further action is warranted.Cook will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THREE-WAY PLASTIC STOPCOCK
Type of Device
KGZ ACCESSORIES, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key10148487
MDR Text Key196425253
Report Number1820334-2020-01119
Device Sequence Number1
Product Code KGZ
UDI-Device Identifier00827002002197
UDI-Public(01)00827002002197(17)240204(10)9496210
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/04/2024
Device Catalogue NumberPTWS-2FLL-MLL-R
Device Lot Number9496210
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/04/2020
Initial Date FDA Received06/12/2020
Supplement Dates Manufacturer Received07/23/2020
Supplement Dates FDA Received08/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-