Information was received on 06/15/2020 indicating that the event occurred during testing and no patient was involved.Device evaluation completion date: 07/01/2020.Device evaluation: one smiths medical cadd solis hpca pump was returned for analysis in a good condition.During analysis, the customers concern regarding failed accuracy was unable to be confirmed.Delivery accuracy testing found the pumps average delivery error to be within the published specification of +/-6%.However, service recommended the expulsor to be replaced as preventive measure.Based on the evidence, the complaint was not confirmed, and the problem source of the reported event is unknown.
|