ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752022
Device Problem Device Slipped (1584)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/19/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported when the infusion cannula was inserted into the trocar, it would fall down all the time without being self-standing during a procedure.The procedure was completed while supporting it with his finger until the end (without product replacement).There was no harm to the patient.
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.Only the used infusion cannula line was returned.It was noted the trocar was returned for evaluation.The infusion cannula tip was measured by inserting the 25 gauge (ga) into the go-no-go gauge and confirmed the cannula tip to be 25ga.The infusion cannula tip was then inserted into a 25ga trocar and could fully engage together with no anomalies observed.It was confirmed 25ga trocars and 25ga infusion cannula was built into this finished goods lot.A review of the lot history indicates this issue to be an isolated event for this finished goods lot.The root cause of the customer's complaint could not be established as the returned met specifications.After investigation of this complaint, it has been determined that this sample met specifications; therefore, no corrective action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key10155472
MDR Text Key195538307
Report Number1644019-2020-00295
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 07/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Catalogue Number8065752022
Device Lot Number2312363H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2020
Initial Date Manufacturer Received 06/10/2020
Initial Date FDA Received06/15/2020
Supplement Dates Manufacturer Received07/22/2020
Supplement Dates FDA Received07/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CONSTELLATION VISION SYSTEM
-
-