• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, LARGE, 12, STERILE, SINGLE USE; HF-RESECTION ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, LARGE, 12, STERILE, SINGLE USE; HF-RESECTION ELECTRODES Back to Search Results
Model Number WA22503D
Device Problems Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/02/2020
Event Type  malfunction  
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during a therapeutic transurethral resection of the prostate (turp) procedure, the loop wire at the distal end of the hf resection electrode broke off and fell into the patient.It is unknown whether or not a fragment remained inside the patient.No further information was provided but there was no report about an adverse event or patient injury.
 
Manufacturer Narrative
Corrected data: d4 - lot number: 1000042837.D4 - expiration date: 14.04.2024.H4 - device manufacturer date: 10.06.2019.Device evaluation: the suspect medical device was returned to the manufacturer for investigation.The evaluation/investigation confirmed that the hf resection electrode shows a broken loop wire with melted points at the ends.The cause of this damage is most likely wear and tear.Therefore, this event/incident was attributed to use error.Furthermore, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the hf resection electrode without showing any abnormalities.The case will be closed from olympus side with no further actions but the reported event/incident will be recorded for trending and surveillance purposes.Furthermore, the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, LARGE, 12, STERILE, SINGLE USE
Type of Device
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key10161928
MDR Text Key198325611
Report Number9610773-2020-00135
Device Sequence Number1
Product Code FAS
UDI-Device Identifier14042761068205
UDI-Public14042761068205
Combination Product (y/n)N
PMA/PMN Number
K120418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 07/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/14/2024
Device Model NumberWA22503D
Device Catalogue NumberWA22503D
Device Lot Number1000042837
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/05/2020
Initial Date FDA Received06/17/2020
Supplement Dates Manufacturer Received07/27/2020
Supplement Dates FDA Received07/31/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-