Model Number C315HIS02
Device Problem Mechanical Problem (1384)
Patient Problem No Code Available (3191)
Event Date 05/29/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during use of the guide catheter the valve was loose and the amount of reverse blood was considerable.The guide catheter was replaced and the procedure was completed.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the delivery catheter was returned and analyzed.The analysis indicated that the septum of the catheter was damaged.Blood was observed on the septum of the delivery catheter.Visual analysis of the delivery catheter indicated damage during use.The analyst noted the delivery catheter was returned without the dilator.There was a septal cap on the infusion port on the hub of the delivery catheter.There was a tear in the septum in the hub of the delivery catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEFLECT
Type of Device
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key10162753
MDR Text Key195432738
Report Number9612164-2020-02237
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
PMA/PMN Number
K101885
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC315HIS02
Device Catalogue NumberC315HIS02
Device Lot Number0010181159
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/03/2020
Initial Date Manufacturer Received 05/29/2020
Initial Date FDA Received06/17/2020
Supplement Dates Manufacturer Received06/23/2020
Supplement Dates FDA Received06/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-