• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN UNWTD KANG FEED TUBE 14FR; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN UNWTD KANG FEED TUBE 14FR; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884711519
Device Problem Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/04/2020
Event Type  malfunction  
Manufacturer Narrative
The complainant indicated that the device will not be returned for evaluation; therefore, a failure analysis is not available, and we are not able to determine the relationship between this device and the cause for this event.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.If additional information or the sample is received, the investigation will be reopened and responded to accordingly.
 
Event Description
The customer reported that on (b)(6) 2020, they inserted 1st silicone ngt tube and noticed milk had a very slow flow.On (b)(6) 2020, the ngt tube was removed and noticed that there was a kink at the ngt tube near to the ending part.New silicone ngt tube was inserted.There was no patient injury.
 
Manufacturer Narrative
The lot number was provided and the device history record (dhr) was reviewed indicating that the product was released accomplishing all quality standards.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.One photograph was received for the analysis.Based on the photograph provided, it is not possible to confirm the reported condition.The customer has confirmed that the device was discarded.A physical sample is required to evaluate the reported issue.Without a physical sample being returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the issue and root cause analysis.A corrective action is not applicable at this time.The current process is running according to product specifications, meeting all quality acceptance criteria.We will keep monitoring the process for any adverse trends that require immediate attention.This complaint will be used for qa tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNWTD KANG FEED TUBE 14FR
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
COVIDIEN
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX  22500
MDR Report Key10165868
MDR Text Key198740184
Report Number9612030-2020-02502
Device Sequence Number1
Product Code KNT
UDI-Device Identifier10884521009691
UDI-Public10884521009691
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 08/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number8884711519
Device Catalogue Number8884711519
Device Lot Number1916507864
Was Device Available for Evaluation? No
Date Manufacturer Received06/03/2020
Patient Sequence Number1
Patient Age78 YR
-
-