• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. PROFORE KIT LATEX FREE; BANDAGE, ELASTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. PROFORE KIT LATEX FREE; BANDAGE, ELASTIC Back to Search Results
Model Number 66020626
Device Problem Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/22/2020
Event Type  malfunction  
Event Description
It was reported that during set up the coband layer were found to stick together and very difficult to pull apart.The product was not used on the patient and a back-up was available to continue the treatment.There was no delay and no patient harm was reported.
 
Manufacturer Narrative
The device that was intended for use in treatment was not returned for evaluation with no additional information provided we have not been able to establish a relationship between the device and the reported event or determine a root cause.A lot number was not provided a document review was not performed.A complaint history review found other related failures.Probable cause may be a manufacturing related issue.Smith and nephew will continue to monitor for any adverse trends relating to this product.No further investigation is required.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, has not been returned for evaluation.The information provided has been reviewed and we cannot confirm a relationship between the device and the reported event.No lot/serial number has been provided, therefore a review of device history is not possible.A further review of the manufacturing process was also performed, and a root cause of inadequate standard operating procedure has been assigned.A complaint history review was performed for the product and event description, there have been further instances of this nature.Corrective action has been assigned regarding this event to reduce the probability of further reoccurrences.This investigation is now complete, smith + nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROFORE KIT LATEX FREE
Type of Device
BANDAGE, ELASTIC
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key10171835
MDR Text Key195763543
Report Number8043484-2020-00844
Device Sequence Number1
Product Code FQM
UDI-Device Identifier05000223496401
UDI-Public5000223496401
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number66020626
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/08/2020
Initial Date FDA Received06/18/2020
Supplement Dates Manufacturer Received05/08/2020
11/17/2021
Supplement Dates FDA Received03/15/2021
11/30/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-