RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Failure to Charge (1085); Battery Problem (2885)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 18jun2020.
|
|
Event Description
|
The customer reported the battery failed.The customer indicated the unit powered on with out alternating current (ac) came up battery fail.The customer noted that the battery was down to 10 volts, swapped battery and it did not charge.The customer verified that the output voltage from the v60 is 14 volts and the battery was 2 years old.The voltage on the battery with the volt meter reads 0 volts.Swapped in another battery that displays 9.89 volts which this battery has been on shelf for 2 years.The customer has battery that the date reads 2017 and it will not charge either.Verified battery failed error code in the significant event log.The remote manufacture service technician advised the customer to replace the battery.The device did not have patient involvement at the time the issue was discovered, therefore, there was no patient or user harm reported.
|
|
Manufacturer Narrative
|
G4:04dec2020.B4:05dec2020.The customer confirmed the issue was resolved by replacing the battery.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|