• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II UNI RESECT AP BLOCK LM/RL SZ 1; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II UNI RESECT AP BLOCK LM/RL SZ 1; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 74016403
Device Problem Break (1069)
Patient Problem Injury (2348)
Event Date 02/04/2020
Event Type  Injury  
Event Description
It was reported that during use when the instruments were inside the patient, the cut blocks produced metal shaving when contacted by the drill in the femur prep process.There was a delay between 0-30 min.The surgery was finished with the same device.
 
Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device does not confirm the stated failure mode.The device was various nicks and burrs.A dimensional evaluation confirmed the stated failure mode.The device was found to be out of specification.A health hazard evaluation was also performed and concluded that the risk level was determined to be low.The clinical/medical evaluation concluded that this case reports that while using the journey ii cutting blocks metal shavings were produced when drilling.It is unknown if the metal shavings were removed from the patient.Without the requested clinical information, the reported issue could not be further assessed.The material for the device is 17-4 ph stainless steel and is manufactured and intended as an externally communicating device, not approved for implantation; therefore, long-term implantation data is not available.The patient impact beyond possible micro-motion and/or migration, and local irritation/discomfort cannot be determined.No further medical assessment can be rendered at this time.Should additional clinical documentation become available in the future, the clinical/medical task may be re-evaluated.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.A potential probable cause statement could be but is not limited to an error in manufacturing.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II UNI RESECT AP BLOCK LM/RL SZ 1
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10174765
MDR Text Key195760412
Report Number1020279-2020-02609
Device Sequence Number1
Product Code HSX
Combination Product (y/n)N
PMA/PMN Number
K190085
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number74016403
Device Lot Number19GM01454
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/01/2020
Distributor Facility Aware Date02/04/2020
Initial Date Manufacturer Received 02/04/2020
Initial Date FDA Received06/19/2020
Supplement Dates Manufacturer Received09/08/2021
Supplement Dates FDA Received09/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-