• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100 A
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Patient Involvement (2645)
Event Date 05/21/2020
Event Type  malfunction  
Manufacturer Narrative
Vyaire medical file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.No root cause has been determined yet.The customer reported that they checked the pressure sense line with no water or residual inside.They also disconnected and checked the p3 related to the frequency potentiometer as well as the regulators on the plc.For now, they have reached the maximum frequency desired but they will be monitoring this case.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information was received.
 
Event Description
The customer reported that the 3100 a unit's map fluctuates 2 cm with a maximum frequency of 8.The customer confirmed that there was no patient involvement associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa, il
Manufacturer (Section G)
VYAIRE MEDICAL
1100 bird center drive
palm springs, ca
Manufacturer Contact
mindy faber
26125 n. riverwoods blvd.
mettawa, il 
7570116
MDR Report Key10177097
MDR Text Key196212971
Report Number2021710-2020-12130
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003055
UDI-Public(01)10846446003055(11)20040629
Combination Product (y/n)N
Reporter Country CodeLE
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number766522
Was Device Available for Evaluation? Yes
Date Manufacturer Received05/21/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/29/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-