Qn#(b)(4).A device history record review was performed on the epidural catheter with no relevant findings.The customer reported a hole was discovered in the catheter during use.The customer returned one snaplock assembly, one epidural catheter, and lidstock.The returned components were received connected together (reference attached files inp1900077481).The returned sample was visually examined with and without magnification.Visual examination of the returned snaplock assembly revealed the snaplock appears typical with no observed defects or anomalies.Visual examination of the returned catheter revealed that the catheter appears used as biological material can be seen on the inner coils.Microscopic examination of the catheter revealed the catheter's extrusion appears to have a hole at approximately 28.1mm (10171599) from the proximal end.No other damage was observed.Functional inspection was performed on the returned sample.A functional leak test was performed per amrq-000017 section 7.5 rev.7 using the returned catheter and returned snaplock assembly with the lab leak tester (ref-002902).The proximal end of the catheter was inserted into the snaplock assembly until it bottomed out and the components were locked.The catheter was confirmed to be secured by tugging gently.The snaplock assembly was then connected to the lab leak tester and the pressure was increased to 10 psi to establish flow.The distal end of the catheter was then capped off and the pressure was increased to 25 psi for 30 seconds.A leak was detected coming from the same location where the catheter has a hole at 28.1mm from the proximal end, which was revealed during the visual inspection.No other leaks were detected.A corrective action is not required at this time as the investigation shows no evidence to suggest a manufacturing related cause.All epidural catheters are tested for leaks at the time of manufacturing.A device history record review performed on the epidural catheter showed no evidence to suggest a manufacturing related cause.The hole in the extrusion was detected during use.Therefore, based on the condition of the sample received and the time of discovery indicate unintentional user error caused or contributed to this event.The reported complaint of the catheter having a hole in the extrusion was confirmed based on the sample received.During the functional inspection, the returned epidural catheter was confirmed to leak from where the catheter had a hole in the extrusion at approximately 28.1mm from the proximal end.All epidural catheters are 100% tested for leaks at the time of manufacturing.A device history record review was performed on the epidural catheter with no evidence to suggest a manufacturing related issue.The hole in the extrusion was detected during use.Therefore, based on the time of discovery and the condition of the sample received, unintentional user error caused or contributed to this event.
|