Catalog Number V041079000000 |
Device Problems
Detachment of Device or Device Component (2907); Appropriate Term/Code Not Available (3191)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
There has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.Please note that while this product is not sold in the us, it is considered similar to products that are marketed in the us by dentsply sirona.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
|
Event Description
|
In this event it was reported that a contra angle 6:1 for vdw.Gold/silver would not hold files.The event outcome is unknown as of this mdr evaluation.
|
|
Manufacturer Narrative
|
Additional information received indicating that no injury resulted.Contra angle 101995 (2017) (collet defective, drive parts are very dirty) defective, cleaned.Note: the covid-19 pandemic caused a disruption in normal business activities, resulting in late submission of this report.
|
|