• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JUNI 2 IN 1 FEM BLK RM SZ 3; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JUNI 2 IN 1 FEM BLK RM SZ 3; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 74013463
Device Problem Material Protrusion/Extrusion (2979)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/27/2020
Event Type  malfunction  
Event Description
It was reported that rust was found in the cutting block before the surgery started.The tech sent the case to safety processing and infectious control.Hospital has requested to replace the item affected and will not process trays until replaced.Procedure was completed without delay with the same device.
 
Manufacturer Narrative
The device, intended use in treatment, was not returned for evaluation.The photo provided confirmed the juni 2 in 1 fem blk rm sz 3 shows discoloration.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUNI 2 IN 1 FEM BLK RM SZ 3
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10183435
MDR Text Key196064035
Report Number1020279-2020-02656
Device Sequence Number1
Product Code HSX
UDI-Device Identifier03596010630049
UDI-Public03596010630049
Combination Product (y/n)N
PMA/PMN Number
K113038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74013463
Device Catalogue Number74013463
Initial Date Manufacturer Received 05/28/2020
Initial Date FDA Received06/23/2020
Supplement Dates Manufacturer Received12/15/2020
Supplement Dates FDA Received12/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-