• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SHUNT SENSOR SYS500; BLOOD GAS MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SHUNT SENSOR SYS500; BLOOD GAS MONITOR Back to Search Results
Model Number CDI510H
Device Problem Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 06/03/2020
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during setup the shunt sensor was found to be wet specifically in the area where it was attached to the blood parameter monitor (bpm) in the artery side.No patient involvement as this occurred during setup.Product was changed out.Procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on june 23, 2020.Upon further investigation of the reported event, the following information is new and/or changed: d10 (device availability - added date returned to manufacturer); g4 (date received by manufacturer) ; g7 (indication that this is a follow-up report); h2 (follow-up due to additional information) ; h3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.).All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: h6 (identification of evaluation codes 10, 11, 3331, 213, 67).Method code #1: 10 - testing of actual/suspected device, method code #2: 11 - testing of device from same lot/batch retained by manufacturer, method code #3: 3331 - analysis of production records, results code: 213 - no device problem found, conclusions code: 67 - no problem detected.The returned sample was visually inspected upon receipt, noted that the unit did not have any buffer solution in it.It was leak tested by connecting with the calibrated manometer, submerged into a water bath, and pressurized up to 1030 mmhg, and no leaks were noted.The retention sample from the same lot number was visually inspected and confirmed to have no traces of buffer on the outside of the unit or inside the pouch.The retention sample was then pressurized with air up to 1030 mmhg, submerged in a water bath, and observed for any leaks.No leaks were noted on the retention sample.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHUNT SENSOR SYS500
Type of Device
BLOOD GAS MONITOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key10184143
MDR Text Key196089339
Report Number1124841-2020-00144
Device Sequence Number1
Product Code DRY
UDI-Device Identifier00699753160767
UDI-Public(01)00699753160767
Combination Product (y/n)N
PMA/PMN Number
K972962
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2020
Device Model NumberCDI510H
Device Catalogue NumberN/A
Device Lot NumberYA14
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/03/2020
Initial Date FDA Received06/23/2020
Supplement Dates Manufacturer Received06/25/2020
08/07/2020
Supplement Dates FDA Received07/16/2020
08/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-