It was reported that the device delivered several inappropriate shocks due to noise and oversensing on the right ventricular (rv) channel.There was also noise present on the left ventricular (lv) and right atrial (ra) channels.Technical support was contacted, and there was also high impedance noted in all three channels as well.The device was explanted and replaced, and during the procedure, there was bodily fluid noted in the device header.The patient was stable following the procedure.
|
The reported events of noise, oversensing, inappropriate therapy, and high defibrillation impedance were confirmed via reviewing of the device image.Noise was seen on both of the stored egms and on real-time egm.The device was tested in the laboratory, and noise was also observed during testing.It is believed noise was caused by an unregulated internal supply voltage.The unregulated supply voltage was caused by an intermittent capacitor¿s connection in the hybrid.As a result of this finding, abbott is performing further investigation for connection anomaly with hybrid capacitor c10.Additional findings observed during analysis that are unrelated to the reported event.Premature battery depletion was confirmed by analysis.A device evaluation was performed, and no sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
|