Model Number 71440001
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 05/29/2020
Event Type  malfunction  
Event Description
It was reported that upon inspection noticed that the small knob was broken off.No case involved.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned genesis ii spc fixed housing collet confirms the knob broke off.The broken piece was returned.The device was manufactured in 2009.The device exhibits signs of extensive use and wear.A complaint history review found related failures; this failure mode will be trended by post market surveillance to assess for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII SPC P/S HOUS COLLET S 3-8
Type of Device
PRSTHSS, KN, PTLL/FMRTBL, SM-CNSTRND, NCMNTD, PRS, CTD, PLYMR/MTL/PLYMR
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10188617
MDR Text Key196224799
Report Number1020279-2020-02696
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00885556033401
UDI-Public00885556033401
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 07/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440001
Device Catalogue Number71440001
Device Lot Number09HM02704
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2020
Date Manufacturer Received07/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-