SMITH & NEPHEW, INC. GII SPC P/S HOUS COLLET S 3-8; PRSTHSS, KN, PTLL/FMRTBL, SM-CNSTRND, NCMNTD, PRS, CTD, PLYMR/MTL/PLYMR
|
Back to Search Results |
|
Model Number 71440001 |
Device Problem
Break (1069)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/29/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that upon inspection noticed that the small knob was broken off.No case involved.
|
|
Manufacturer Narrative
|
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned genesis ii spc fixed housing collet confirms the knob broke off.The broken piece was returned.The device was manufactured in 2009.The device exhibits signs of extensive use and wear.A complaint history review found related failures; this failure mode will be trended by post market surveillance to assess for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|