HAEMONETICS CORPORATION CELL SAVER ELITE AUTOTRANSFUSION SYSTEM; CELL SAVER ELITE SET - 125ML Back to Search Results
Model Number CSE-P-125
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/16/2020
Event Type  malfunction  
Manufacturer Narrative
No patient information was provided at this time.The bowl from the cell saver elite set was returned and evaluated by haemonetics.It was noted during investigation that blood was seen on the inner core.Bowl decompression testing confirmed the blood outflow from the inner core and the outer core welding.
 
Event Description
On january 17 2020 haemonetics was notified of a long empty error message which was displayed on the 3rd cycle of the return process during a cardiac procedure in japan, utilizing the cell saver® elite® autotransfusion system and cell saver® elite set - 125ml.A total blood volume of 722ml was processed and there was no reported impact to patients' health.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL SAVER ELITE AUTOTRANSFUSION SYSTEM
Type of Device
CELL SAVER ELITE SET - 125ML
Manufacturer (Section D)
HAEMONETICS CORPORATION
125 summer street
boston, ma
Manufacturer (Section G)
HAEMONETICS CORPORATION
125 summer street
boston, ma
Manufacturer Contact
janice kiser
125 summer street
boston, ma 
MDR Report Key10189195
MDR Text Key196972268
Report Number1219343-2020-00050
Device Sequence Number1
Product Code CAC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120586
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2022
Device Model NumberCSE-P-125
Device Lot Number0519040
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/17/2020
Initial Date FDA Received06/24/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/26/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-