• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. DISPOSABLE FIRSTPASS SUTURE PASSER, SELF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. DISPOSABLE FIRSTPASS SUTURE PASSER, SELF Back to Search Results
Catalog Number 22-4038
Device Problems Mechanical Jam (2983); Ejection Problem (4009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/04/2020
Event Type  Injury  
Event Description
It was reported that the shoulder rotator cuff procedure, the jaw of the firstpass could not be opened and the trigger was abnormal.The procedure was completed by using a epidural needle and no significant delay was reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10: h2, h3, h6.The reported device, used in treatment, was returned for evaluation.There was a relationship found between the reported incident and the returned device.A review of the device history records for the reported lot number showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.A review of the instructions for use found: - do not use this device as a lever for manipulating hard tissue or bone.Excessive force should not be applied to the device when manipulating soft tissue, bone, or hard objects.Misuse of this device may result in bent or broken distal tip or jaws.- when passing suture multiple times always check the tip of the device for damage or debris between passes.Clear any visible debris between passes.Discontinue use if the device becomes damaged between passes.Clinical/medical evaluation was completed and concluded that no relevant supporting clinical information has been provided to assist with a clinical investigation, and the patient's current condition is unknown.Therefore based on insufficient information, a thorough clinical assessment cannot be performed at this time.Should any additional clinical information be provided this complaint will be re-evaluated.Visual inspection shows no manufacturing abnormalities on the device.The instrument was returned with a closed top bracket which could not be opened.During functional evaluation the first step of the trigger performed as specified; pressing the lever the needle was extracted as intended; however the top bracket remains closed.The complaint was confirmed.Factors that could have contributed to the reported event include: (1) excessive force (2) tissue thickness.(3) damage or debris in the device tip between passes.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE FIRSTPASS SUTURE PASSER, SELF
Type of Device
PASSER
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key10190058
MDR Text Key196266262
Report Number3006524618-2020-00415
Device Sequence Number1
Product Code HWQ
UDI-Device Identifier00885556724545
UDI-Public885556724545
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/03/2022
Device Catalogue Number22-4038
Device Lot Number2032521
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/04/2020
Initial Date FDA Received06/24/2020
Supplement Dates Manufacturer Received06/04/2020
Supplement Dates FDA Received09/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-