MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION COMET; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION COMET; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number 8900
Device Problem Calibration Problem (2890)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/17/2020
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: the product was returned to boston scientific for analysis.Returned product consisted of an ffr comet pressure wire connected to the occ cable.The tip, device shaft, sensor port and the coefficient values were examined for damage or any irregularities.The wire showed 1 kink located 163.5cm from the tip.The tip showed bend damage.There was evidence of peeled coating at the 163.5cm location.The occ handle was connected to the ffr link to verify the signal strength.The signal was present and showed green lights as designed.The occ handle was then connected to the bench top testing equipment and the wire was inserted into the pressure chamber.The pressure was increased to verify the sensor was indeed reacting to the pressure increases and decreases.The pressure sensor functioned as designed.The coefficient values were confirmed to be programmed.The occ handle was again connected to the ffr link.The device was then connected to the polaris (ilab) test equipment via bluetooth signal.The wire communicated to the polaris system and zeroed as designed.With the wire inserted into the test pressure chamber, the wire transferred a pressure waveform to the polaris which indicates a functioning wire.The wire was removed from the occ handle with no issues.The sensor port showed no residue of body fluids.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The complaint was not confirmed for signal/equalizing issues.
 
Event Description
It was reported that a comet wire could not equalize.During preparation and while outside the patient, a comet pressure guidewire could not be equalized with a polaris cart.It was noted that no sedation was given.The device was removed to complete the procedure.No patient complications were reported in relation to this event and the patient was reported to be stable following the procedure.However, device analysis revealed peeled coating.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMET
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10191772
MDR Text Key196327648
Report Number2134265-2020-07723
Device Sequence Number1
Product Code DXO
UDI-Device Identifier08714729904397
UDI-Public08714729904397
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K151610
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/17/2022
Device Model Number8900
Device Catalogue Number8900
Device Lot Number0025226215
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/02/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/18/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age69 YR
Patient Weight84
-
-