• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX LUBRICATH FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX LUBRICATH FOLEY CATHETER Back to Search Results
Catalog Number 0165L14
Device Problems Leak/Splash (1354); Decrease in Pressure (1490)
Patient Problem No Patient Involvement (2645)
Event Date 06/04/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the balloon had a leak prior to insertion.
 
Event Description
It was reported that the balloon had a leak prior to insertion.
 
Manufacturer Narrative
The reported event was confirmed as manufacturing-related.Sample was evaluated and observed there was pinhole on sac collar of catheter that allowed water to leak out.Pinhole was examined under microscope and noted to be short and smooth.Exact cause of sample condition have been determined and confirmed as a manufacturing related process due to poor dipping process.The device history record was reviewed and found a possible manufacturing issue that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.¿ use luer tip syringe to inflate with stated ml of sterile water.Or ¿ for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.For urological use only.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Do not resterilize single use only.Do not reuse.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Consult instructions for use.Do not use if package is damaged." correction: d4.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX LUBRICATH FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key10199868
MDR Text Key197107313
Report Number1018233-2020-04105
Device Sequence Number1
Product Code EZC
Combination Product (y/n)N
PMA/PMN Number
K910846
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0165L14
Device Lot Number9NR142
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2020
Initial Date Manufacturer Received 06/04/2020
Initial Date FDA Received06/26/2020
Supplement Dates Manufacturer Received09/23/2020
Supplement Dates FDA Received10/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-