LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-02-50
Device Problem Pumping Problem (3016)
Patient Problem No Patient Involvement (2645)
Event Date 05/29/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a centrifugal pump system with tubing clamp was giving an error code associated to a defective shaft angle encoder during setup.There was no patient involvement.
 
Manufacturer Narrative
H.10: through follow-up communication livanova learned that the user cycled power on the panel when the reported issue occurred, however, this did not solve the issue.A livanova field service representative was dispatched to the facility and found out the the biomedical engineers of the hospital replaced the control panel on their own.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key10200071
MDR Text Key197283156
Report Number9611109-2020-00379
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-50
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/29/2020
Initial Date FDA Received06/26/2020
Supplement Dates Manufacturer Received06/18/2020
Supplement Dates FDA Received07/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-