Model Number N/A |
Device Problems
Pumping Stopped (1503); Temperature Problem (3022)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/04/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The production device history record (dhr) for this iabp unit is not required to be reviewed per getinge standard operating procedure since the device manufacture date is greater than one year from the aware date.A getinge field service engineer (fse) evaluated the iabp and was unable to reproduce the reported issue.The customer's biomed ran the iabp overnight on a test balloon at 120 bpm and the iabp displayed no over-temp messages.The fse then performed all calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
|
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) displayed high temperature error (system over temperature message) on the screen and stopped pumping.The iabp was replaced with another iabp and therapy continued.There was no harm or injury to the patient and no adverse event was reported.
|
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) displayed high temperature error (system over temperature message) on the screen and stopped pumping.The iabp was replaced with another iabp and therapy continued.There was no harm or injury to the patient and no adverse event was reported.
|
|