• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problems Pumping Stopped (1503); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/04/2020
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this iabp unit is not required to be reviewed per getinge standard operating procedure since the device manufacture date is greater than one year from the aware date.A getinge field service engineer (fse) evaluated the iabp and was unable to reproduce the reported issue.The customer's biomed ran the iabp overnight on a test balloon at 120 bpm and the iabp displayed no over-temp messages.The fse then performed all calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) displayed high temperature error (system over temperature message) on the screen and stopped pumping.The iabp was replaced with another iabp and therapy continued.There was no harm or injury to the patient and no adverse event was reported.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) displayed high temperature error (system over temperature message) on the screen and stopped pumping.The iabp was replaced with another iabp and therapy continued.There was no harm or injury to the patient and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
MDR Report Key10203240
MDR Text Key196750609
Report Number2249723-2020-00989
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/04/2020
Initial Date FDA Received06/26/2020
Supplement Dates Manufacturer Received07/08/2020
Supplement Dates FDA Received07/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-