BOSTON SCIENTIFIC CORPORATION CAPTIFLEX; SNARE, FLEXIBLE Back to Search Results
Model Number M00562402
Device Problems Entrapment of Device (1212); Failure to Cut (2587)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 06/03/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that a captiflex medium oval flexible snare was used in the colon during a colonoscopy procedure performed on (b)(6), 2020.According to the complainant, during the procedure, the snare was not cutting correctly.Reportedly, the target polyp needed to be ripped off or torn off causing patient bleeding.It was reported that the bleeding resolved on its own or by clipping and that the patient did not have any history of coagulopathy or bleeding disorders and was not on any anticoagulants.Also, the snare was securely attached to the active cord and no visible issues were noted with the cautery pin.The procedure was completed with another different captiflex snare.The patient condition following the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTIFLEX
Type of Device
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key10209451
MDR Text Key197294188
Report Number3005099803-2020-02499
Device Sequence Number1
Product Code FDI
UDI-Device Identifier08714729501640
UDI-Public08714729501640
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131700
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 06/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2023
Device Model NumberM00562402
Device Catalogue Number6240-40
Device Lot Number0025111007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/03/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/29/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-