(b)(6).Device is a combination product.Device evaluated by mfr.: promus element,mr,ous 2.75x38mm stent delivery system was returned for analysis.Device returned with mandrel and stent protector attached.Removed without issue.A visual examination of the stent found stent damage.Stent strut damage in midsection of stent, struts lifted.The undamaged stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed signs of distal tip damage.A visual and tactile examination of the hypotube shaft found multiple kinks as well as a break at the distal end of the strain relief.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during analysis.
|
Reportable based on device analysis completed on 10jun2020.It was reported that shaft damage occurred.Vascular access was obtained via the right femoral artery.The 80% stenosed, 32mm x 2.75mm, concentric, de novo target lesion with a bend of >45 and <90 degrees was located in severely tortuous and moderately calcified right coronary artery (rca).After a 6f jr non-bsc guide catheter and a non-bsc guide wire crossed the lesion, predilation was performed with a 2x12mm maverick balloon catheter resulting to 40% residual stenosis.A 2.75x38mm promus element drug-eluting stent was advanced for treatment.However, when the physician was trying to take the device in a challenging rca, the hypotube was damaged and it could be not moved forward.The device was removed and the procedure was completed with another of the same device.No patient complications were reported and the result was good with good timi 3 flow.The patient was moved to the intensive care unit safely.However, returned device analysis revealed stent damaged.
|