• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERATHON MEDICAL ULC GLIDESCOPE CORE SMART CABLE; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERATHON MEDICAL ULC GLIDESCOPE CORE SMART CABLE; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 0800-0602
Device Problems Loose or Intermittent Connection (1371); Poor Quality Image (1408)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/29/2020
Event Type  malfunction  
Manufacturer Narrative
The gs core 10-inch monitor and the gs video baton 2.0 large were returned to verathon for evaluation.A verathon technical service representative evaluated the returned gs core 10-inch monitor and the gs video baton 2.0 large devices and no video issues were found; the fault could not be reproduced.The units functioned as intended.A damaged camera housing was encountered.The gs core smart cable was not returned for analysis.The gs core 10-inch monitor was returned to the customer and the gs video baton 2.0 large was scrapped and a replacement sent to the customer.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Event Description
The customer reported that during a patient procedure, using a glidescope core system (gs core 10-inch monitor, gs video baton 2.0 large and gs core smart cable) the image froze.The customer's biomed also reported that the image was splitting.The biomed stated that both issues seemed to occur when the connection between the cable and the baton became loose.No delay in the procedure, use of a backup device, or harm to the patient or user was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE CORE SMART CABLE
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, wa 
6295760
MDR Report Key10219648
MDR Text Key197850773
Report Number9615393-2020-00139
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 06/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0800-0602
Device Catalogue Number0800-0602
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/04/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-