The reported event was confirmed, however the cause was unknown.The device used for the treatment.The device was affected by the reported failure.Visual evaluation of the returned sample noted that one slightly opened (in original packaging), an unused bulb irrigation syringe was received.The strip packaging of the syringe was noted to have an 0.7795" long opening near the tip of the syringe.The packaging appears as though it was sealed at one point.Although the reported event was confirmed, the root cause could not be determined.A potential root cause for this failure mode could be due to an incorrect parameter of sealing.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "this is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Single use.Do not resterilize.Do not use if package is damaged.Latex-free.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.Caution: federal (usa) law restricts this device to sale by or on the order of a physician." h11:section a through f the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|