Model Number 1400
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Device evaluation: one smiths medical cadd-legacy duodopa ambulatory infusion pump was returned for analysis in used condition.Delivery accuracy tests were performed and the pump was found to be delivering properly and within specification.The pump passed all tests and was found to be operating properly.Based on the investigation, the complaint allegation was not confirmed.
 
Event Description
Information was received indicating that a smiths medical cadd-legacy duodopa ambulatory infusion pump was "not delivering the correct amount of gel." information was not available on the specific nature of the delivery problem.Per reporter the pump was subsequently exchanged.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-LEGACY DUODOPA AMBULATORY INFUSION PUMP
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis, mn
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
3350 granada ave. n.
suite 100
oakdale, mn
Manufacturer Contact
dave halverson
6000 nathan lane north
minneapolis, mn 
3833310
MDR Report Key10220659
MDR Text Key197206733
Report Number3012307300-2020-06477
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 07/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1400
Device Catalogue Number21-1400-03
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2020
Date Manufacturer Received06/03/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/26/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-