Model Number 74018911
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/10/2020
Event Type  malfunction  
Event Description
It was reported that during surgery, while outside the patient, the spring lever on the item snapped.Procedure was concluded with a backup device from smith and nephew.No delay and no injury reported.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection confirmed the spring is broken off the journey art isrt assm tool and has not been returned.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY ART ISRT ASSM TOOL
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10221092
MDR Text Key197273197
Report Number1020279-2020-02866
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010557292
UDI-Public03596010557292
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 07/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74018911
Device Catalogue Number74018911
Device Lot Number13GPG0017
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-