The initial reporter stated they received a discrepant result for one patient sample tested with the elecsys anti-sars-cov-2 reagent lot number 495464 on a cobas e 411 immunoassay analyzer.No incorrect results were reported outside of the laboratory.The sample was tested using elecsys anti-sars-cov-2 reagent lot number 494813, resulting with a value of 1.42 coi (reactive).The sample was tested using elecsys anti-sars-cov-2 reagent lot number 496298, resulting with a value of 1.30 coi (reactive).The sample was tested using elecsys anti-sars-cov-2 reagent lot number 495464, resulting with a value of 0.989 coi (non-reactive).Elecsys anti-sars-cov-2 reagent lot number 495464 was re-calibrated and the sample was repeated five times on (b)(6) 2020, resulting with the following values: 1.21 coi (reactive), 1.21 coi (reactive), 1.21 coi (reactive), 1.23 coi (reactive), and 1.24 coi (reactive).The serial number of the e411 analyzer is (b)(4).
|