The investigation determined that higher than expected vitros amon results were obtained from vitros performance verifier quality control fluids using vitros amon reagent on a vitros 5600 integrated system.A definitive assignable cause of the event was not established.However, the most likely cause of the event is day to day inconsistencies with quality control fluid handling.Acceptable within run amon and alkp precision test results indicate the vitros 5600 system is performing as expected and is not a contributor of the event or the unacceptable quality control within laboratory precision.The historical vitros amon quality control results indicate unacceptable within laboratory precision.However, acceptable within run amon precision test results verified the system performance; therefore, day to day inconsistencies with fluid handling is the likely cause of the observed imprecision and the higher than expected vitros amon results.In addition, ongoing tracking and trending of complaint data has not identified any signals to suggest there is a systemic quality issue with vitros amon slides lot 1018-0253-8819.
|