BOSTON SCIENTIFIC CORPORATION VIGILANT X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number G247 |
Device Problems
High impedance (1291); Noise, Audible (3273)
|
Patient Problem
Muscle Stimulation (1412)
|
Event Date 02/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) exhibited diaphragm stimulation through the left ventricular (lv) lead and an increase in lv pace impedance measurements, measuring between 1600-2000 ohms, which triggered the device to beep to notify the patient to reach out to their following clinic.The device beeped as the programmed lv beep range measurements were set to 2000 ohms.The normal impedance range for this lv lead is 200-3000 ohms.Technical services was consulted and recommended troubleshooting and programming recommendations.The patient was brought into the clinic and reprogramming was performed.The lv measurements were within range.The device remains in service.No adverse patient effects were reported.
|
|