• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-01
Device Problems Mechanical Problem (1384); Device Slipped (1584); Intermittent Communication Failure (4038)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/17/2020
Event Type  malfunction  
Event Description
The recipient is reportedly experiencing loss of lock.External equipment was exchanged, however, the issue did not resolve.Revision surgery will be scheduled.
 
Manufacturer Narrative
Additional information: section d.7.The recipient's device was reportedly explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed tool damage to the bottom cover, as well as cuts to the electrode in the straight portion and also near the array prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.The no lock condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device failed the residual gas analysis test.This device had moisture that exceeded the residual gas analysis test limit.The source of the problem was a feedthru hermeticity issue from one feedthru vendor.A corrective action was implemented.Feedthru assemblies from this vendor are no longer used.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key10237734
MDR Text Key198294036
Report Number3006556115-2020-00541
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup,Followup
Report Date 06/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/01/2006
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received07/06/2020
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Supplement Dates FDA Received09/29/2020
03/30/2021
07/02/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age17 YR
-
-