SMITH & NEPHEW MEDICAL LTD. FOOTSWITCH, MULTI-FUNCTION, VERSAJET II; LAVAGE, JET Back to Search Results
Catalog Number 66800472
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/11/2020
Event Type  malfunction  
Event Description
It was reported that the cable was found broken and console doesn¿t recognize the foot pedal.No case involved.
 
Manufacturer Narrative
The device, intended for use in treatment, has been returned for evaluation, visual inspection reported exposed wire damage, the functional evaluation found the device failed to communicate the transmission for the foot pedal establishing a relationship between the device and the reported events.The root cause identified as component failure due to cable strained, we could not verify the lot number but we have no reason to believe that the product did not meet the specification at the time of manufacture.A complaint history review found other related events.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
 
Manufacturer Narrative
The device intended to be used in treatment has not been returned for evaluation, with no additional information provided we have not been able to establish a relationship between the device and the reported event or determine a root cause on this occasion.Factors that can contribute to the reported event include, damage, or obstruction, the instructions for use provide comprehensive instructions of the operation, use and limitations of the device.A dhr review was not performed because the lot number provided was inaccurate.The complaint history file contains further instances of the reported event.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOOTSWITCH, MULTI-FUNCTION, VERSAJET II
Type of Device
LAVAGE, JET
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10239643
MDR Text Key197791830
Report Number8043484-2020-01722
Device Sequence Number1
Product Code FQH
Combination Product (y/n)N
PMA/PMN Number
K143115
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup,Followup
Report Date 08/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number66800472
Device Lot Number04893
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2021
Date Manufacturer Received08/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-