EDWARDS LIFESCIENCES; REPLACEMENT HEART VALVE Back to Search Results
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Appropriate Term/Code Not Available (3191)
Patient Problem Aortic Valve Stenosis (1717)
Event Date 06/12/2020
Event Type  malfunction  
Manufacturer Narrative
Additional manufacturer narrative: stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explants and encompasses multiple failure modes, including calcification, non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes may occur singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, the nonstructural dysfunction (nsvd) may also play a role in the development of valvular stenosis.The root cause of this event cannot be conclusively determined with the available information.However, the stenosis in this case was most likely impacted by the progression of the patient¿s underlying valvular disease pathology with or without structural valve deterioration and/or nonstructural dysfunction.The subject device is not available for evaluation as it remains implanted in the patient.Although edwards was unable to perform device analysis, this event was most likely due to a progression of the patient¿s underlying valvular disease pathology.The device history record (dhr) was not reviewed as the device serial number was not provided.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event Description
It was reported that a patient with an edwards surgical valve in the aortic position has been placed under consideration for a valve-in-valve procedure after an implant duration of five (5) years, four (4) months due to severe aortic stenosis.
 
Manufacturer Narrative
Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
MDR Report Key10240920
MDR Text Key198696532
Report Number2015691-2020-12434
Device Sequence Number1
Product Code DYE
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 06/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age89 YR
-
-