The patient was undergoing a thrombectomy procedure in the left internal carotid artery (ica) using a penumbra system jet 7 reperfusion catheter (jet7), penumbra system 3max reperfusion catheter (3maxc), and guide catheter.It was noted that the patient¿s anatomy was severely tortuous in the abdomen.During the procedure, the physician experienced resistance while advancing a guidewire, 3maxc, and jet7 through the balloon catheter.Therefore, the guidewire, 3maxc and, jet7 were removed.It was a reported that while flushing the jet7 on the back table, the jet7 expanded approximately 20mm from the distal tip.The physician then advanced a non-penumbra microcatheter through the guide catheter; however, resistance was encountered while advancing through the tortuous vessels.Therefore, the microcatheter and guide catheter were removed.It was also reported that the guide catheter was found to be kinked on the back table.The physician was unable to complete the procedure using other devices and decided to end the procedure at this point.There was no report of an adverse effect to the patient.
|
Results: the distal shaft of the jet7 was stretched approximately 129.5 cm from the hub.Conclusions: evaluation of the returned jet7 confirmed that the distal shaft of the catheter expanded while being flushed.Further evaluation of the device revealed that the distal shaft was stretched.If the jet7 is forcefully manipulated against resistance, damage such as stretching may occur.During the functional test, a demonstration microcatheter could not be advanced through the returned jet7, and the returned jet7 could not be advanced through a demonstration neuron max due to the stretched distal shaft on the returned jet7.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|