LIVANOVA DEUTSCHLAND S5 DOUBLE HEAD PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problem Pumping Problem (3016)
Patient Problem No Patient Involvement (2645)
Event Date 06/10/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).Through follow-up communication livanova learned that the pump did not start when the tubing was installed and that the right motor does not work after the pump was restarted.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 double head pump was giving a motor control error during training.There was no patient involvement.
 
Event Description
See initial report.
 
Manufacturer Narrative
Following the s5 service manual (cp_sem_48-x0-00, revision 02), the field service technician replaced the hms and the hmf boards.Replacement of the boards solved the reported issue.Technical safety check was successfully performed after the replacement and the unit was put back in service.A review of the dhr could not identify any deviations or nonconformities relevant to the issue.Based on all the above facts, it cannot be ruled out that the reported malfunction was due to a defective motor controller board and motor power amplifier board.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key10244065
MDR Text Key197947813
Report Number9611109-2020-00405
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Was the Report Sent to FDA? No
Date Manufacturer Received07/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-