• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES GORE VIABAHN VBX BALLOON EXPANDABLE ENDOPROSTHESIS; ILIAC COVERED STENT, ARTERIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES GORE VIABAHN VBX BALLOON EXPANDABLE ENDOPROSTHESIS; ILIAC COVERED STENT, ARTERIAL Back to Search Results
Model Number BXA085901A
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/22/2020
Event Type  malfunction  
Manufacturer Narrative
Review of the manufacturing records verified that the lot met release requirements.The device was not returned.Consequently, a direct product analysis was not possible.All information has been placed on file for use in tracking and trending.Reference medwatch #2017233-2020-00439 for the 1st device.
 
Event Description
It was reported to gore on (b)(6) 2020 a procedure was being performed using a gore® viabahn® vbx balloon expandable endoprosthesis.The physician reported the patient had tortuous anatomy which he believes caused the stent to fall off the balloon.He then tried another vbx of the same size and that stent came off the balloon as well.Patient disposition is unknown at this time.This medwatch report is for the 2nd reported device.
 
Manufacturer Narrative
Additional manufacturer narrative: the referenced complaint did not have a returned device.The following evaluation is based on information obtained from the event description and communication summary.Stent dislodgement of the vbx device may be affected by device profile and manufacturing crush force.Device profile is 100% verified during the vbx manufacturing process.The device history file was reviewed and no anomalies were identified.Crush force is controlled through machine maintenance and calibration.Machine history files were reviewed and no non-routine maintenance was identified that would contribute to dislodgement.Based on the device evaluation performed, no manufacturing anomalies were identified to which the event could be definitively attributed.
 
Manufacturer Narrative
Additional manufacturing narrative: c1.Name (#1) - cbas® heparin surface; manufacturer/compounder: w.L.Gore & associates, inc.Lot #21046410.Cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE VIABAHN VBX BALLOON EXPANDABLE ENDOPROSTHESIS
Type of Device
ILIAC COVERED STENT, ARTERIAL
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
MDR Report Key10247886
MDR Text Key199052550
Report Number2017233-2020-00465
Device Sequence Number1
Product Code PRL
UDI-Device Identifier00733132637324
UDI-Public00733132637324
Combination Product (y/n)N
PMA/PMN Number
P160021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup,Followup
Report Date 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/09/2022
Device Model NumberBXA085901A
Device Catalogue NumberBXA085901A
Device Lot Number21046410
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received07/08/2020
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Supplement Dates FDA Received07/27/2020
07/27/2020
07/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-