During follow-up, approximately two weeks after the implant procedure, the pacing impedance of the right ventricular (rv) lead was high and out of range.In addition, failure to capture and low sensing was observed.Abbott technical support was contacted and recommended a revision procedure.The patient underwent surgery, the icd was removed and blood was observed in the rv channel of the header.The electrical values of the rv lead were measured with a pacing system analyzer and all electrical values were in range.The device was explanted and replaced to resolve the event.The patient was stable before, during and after the procedure.
|
The device was returned for analysis and after decontamination, a visual inspection was performed and no anomalies were observed.The reported field event of high impedance was verified in the data stored in the device.The device¿s data showed the ventricular pacing lead impedance measurements were high starting on (b)(6) 2020.The device underwent testing including impedance, sensing, pacing, and high voltage shock output and no anomalies were found.All impedances measured in the lab were normal.The cause of the reported events could not be determined.
|