Model Number 24653 |
Device Problem
Entrapment of Device (1212)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/23/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).
|
|
Event Description
|
It was reported that the delivery system became stuck on the wire.The 100% stenosed target lesion was located in the moderately tortuous superficial femoral artery (sfa).A 7x120x130 eluvia drug-eluting vascular stent system was selected for the procedure.The stent was deployed.Upon removal, the delivery system became stuck on the wire.The delivery system and the wire were removed together.There were no patient complications reported and the patient's status was good post procedure.
|
|
Event Description
|
It was reported that the delivery system became stuck on the wire.The 100% stenosed target lesion was located in the moderately tortuous superficial femoral artery (sfa).A 7x120x130 eluvia drug-eluting vascular stent system was selected for the procedure.The stent was deployed.Upon removal, the delivery system became stuck on the wire.The delivery system and the wire were removed together.There were no patient complications reported and the patient was good post procedure.
|
|
Manufacturer Narrative
|
Initial reporter city: (b)(6).Device evaluated by mfr: returned product consisted of an eluvia self-expanding stent system with a 0.014" guidewire stuck in the device.The outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed a kink to the outer sheath at the nosecone.Microscopic examination revealed no additional damages.The device was x-ray and the proximal inner was prolapsed.Inspection of the remainder of the device, revealed no other damage or irregularities.
|
|
Search Alerts/Recalls
|