BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number 72404252
Device Problems Failure to Cycle (1142); Fluid/Blood Leak (1250); Mechanical Problem (1384); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/19/2020
Event Type  Injury  
Event Description
It was reported that the patient experienced unspecified performance issues with his ams device.
 
Manufacturer Narrative
The complaint component was returned and analyzed, and the reported allegation was not confirmed via product analysis.The ams 700 momentary squeeze (ms) pump was visually inspected; no leaks were found.The kink resistant tubing (krt) was worn to filament; no leak was found.Therefore, this wear would not affect the functionality of the device.The pump was functional test and performed within specification.The event cannot be reproduced or substantiated; therefore, no escalation to ncep, capa or scar is required.
 
Event Description
It was reported that the patient underwent an inflatable penile prosthesis (ipp) procedure due to the device was not cycling.Upon removal, a fluid leak in the tubing was identified.Cylinders and pump were removed and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE
Type of Device
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
10700 bren road w
minnetonka MN 55343
MDR Report Key10254355
MDR Text Key198317817
Report Number2183959-2020-02854
Device Sequence Number1
Product Code FHW
Combination Product (y/n)N
PMA/PMN Number
N970012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 08/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/09/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/20/2018
Device Model Number72404252
Device Catalogue Number72404252
Device Lot Number0144021002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2020
Date Manufacturer Received07/13/2020
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age63 YR
-
-