Model Number CI-1601-04 |
Device Problems
Malposition of Device (2616); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/09/2020 |
Event Type
Injury
|
Event Description
|
The recipient reportedly experienced an incorrect electrode position.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
The recipient reportedly experienced difficulty during initial insertion due to osteoporosis.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed silicone damage on the top and bottom covers, and the lumen was damaged prior to receipt.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|