The aquabeam handpiece was returned for investigation.Visual inspection was able to confirm the reported event that the aspiration metal tubes had come detached from the aquabeam handpiece.Further inspection found that the device was manufactured correctly.The results of the investigation are consistent with too much pressure on the tip of the aquabeam handpiece, however, the root cause of the pressure remains undetermined.This is the second occurrence, with the first occurrence on (b)(6) 2019 (b)(4) therefore, no escalation is required or recommended at this time.Complaint trends are reviewed on a monthly basis.Should a trend arise for this failure mode, further actions will be considered.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|