Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Patient Involvement (2645)
Event Date 06/15/2020
Event Type  malfunction  
Manufacturer Narrative
During preventive maintenance at customer site, the autopulse platform (sn (b)(4)) failed initial functional testing and displayed user advisory (ua) 2 (compression tracking error) error message.The root was due to defective load cells, likely due to a defective component or mishandling such as drop.During visual inspection, no issue was observed during functional testing, the platform displayed ua 2 error message due to a defective load cell 1 that was confirmed by load cell characterization test.After replacing the load cell 1 to address ua 2, during further functional testing the platform displayed ua 17 (max motor on time exceeded).The root cause of the ua 17 issue was due to the brake gap being out of specification.The brake gap was readjusted to remedy the fault.The platform was manufacture on 16 july 2016 and there was no preventive maintenance performed.After service, the autopulse platform was subjected to the run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test batteries for 15 minutes without any fault or error.The autopulse platform passed the final testing without any fault or error.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaint reported for autopulse platform with sn (b)(4).
 
Event Description
During preventive maintenance at customer site, the autopulse platform (sn (b)(4)) failed initial functional testing and displayed user advisory (ua) 2 (compression tracking error) error message.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE PLATFORM RESUSCITATION
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, ca 
4192922
MDR Report Key10256078
MDR Text Key198756299
Report Number3010617000-2020-00670
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111001076
UDI-Public00849111001076
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0740-16
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/15/2020
Initial Date FDA Received07/09/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/16/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-