Model Number MODEL 100
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Patient Involvement (2645)
Event Date 06/16/2020
Event Type  malfunction  
Manufacturer Narrative
The report of the autopulse platform (sn (b)(4)) displayed ua 45 (driveshaft not at "home" position after poweron/restart) error message and the user tried to rotate the drive shaft to home position, however, the drive shaft remain stuck and was unable to be rotated to home position was confirmed during functional testing and archive data review.The root cause for the ua45 error message was due to the driveshaft not being at home position due to the encoder drive shaft was difficult to rotate and exhibited binding and resistance.The clutch plate was deburred and the drive shaft was rotated to home position to clear the ua45 error message.During visual inspection, no physical damage was observed.During archive data review, ua 45 error message was observed.During functional testing, the autopulse platform displayed user advisory (ua) 45 (not at "home" position after power-on/restart) error message" upon powering on.The root cause for the ua45 error message was due to the driveshaft not being at home.After clearing the ua45 error message, the autopulse platform performed compressions without any fault or error using lrtf (large resuscitation test fixture).Historical records were reviewed for service information related to the reported complaint and there was no similar history of complaint reported for autopulse with serial number (b)(4).
 
Event Description
During shift check, the autopulse platform (sn (b)(4)) displayed user advisory (ua) 45 (not at "home" position after power-on/restart) error message.The user tried to rotate the drive shaft to home position, however, the drive shaft remain stuck and was unable to be rotated to home position.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE PLATFORM RESUSCITATION
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, ca 
4192922
MDR Report Key10256088
MDR Text Key198783560
Report Number3010617000-2020-00671
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/16/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-