Catalog Number 71430416
Device Problem Material Fragmentation (1261)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/17/2020
Event Type  malfunction  
Event Description
It was reported that during tka procedure the plastic on gii ps hi flex isrt tr s5-6 11 was found to be chipped.No delay.Procedure was finished using a s&n back up device.No injuries reported.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection confirmed the gii ps hi flex isrt tr s5-6 11 is chipped.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS HI FLEX ISRT TR S5-6 11
Type of Device
TRAY, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10260682
MDR Text Key198484606
Report Number1020279-2020-03070
Device Sequence Number1
Product Code FSM
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 08/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/01/2020
Device Catalogue Number71430416
Device Lot Number07DM11695
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/17/2020
Initial Date FDA Received07/10/2020
Supplement Dates Manufacturer Received08/10/2020
Supplement Dates FDA Received08/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-