• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 32MM X 48MM; PRSTHSS,HIP,SEMI-CONSTRAINED,METAL/POLYMER,CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 32MM X 48MM; PRSTHSS,HIP,SEMI-CONSTRAINED,METAL/POLYMER,CEMENTED Back to Search Results
Model Number 71337648
Device Problem Mechanical Problem (1384)
Patient Problems Pain (1994); Injury (2348)
Event Date 06/16/2020
Event Type  Injury  
Event Description
It was reported that a revision surgery had to be performed due to pain, swelling and dislocation.This case is not caused by s+n devices defect.Surgeon does not fault the devices.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that this complaint from japan reports a revision secondary to ¿pain, swelling and dislocation¿ it has been communicated that no further information would be forthcoming.Smith and nephew has not received adequate materials (operative reports) to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode and potential harm was previously identified.Possible causes could include but not limited to traumatic injury, patient anatomy or abnormal loading of limb.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 20 DEG XLPE ACET LNR 32MM X 48MM
Type of Device
PRSTHSS,HIP,SEMI-CONSTRAINED,METAL/POLYMER,CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10261755
MDR Text Key198486382
Report Number1020279-2020-03098
Device Sequence Number1
Product Code JDI
UDI-Device Identifier03596010598479
UDI-Public03596010598479
Combination Product (y/n)N
PMA/PMN Number
K113848
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71337648
Device Catalogue Number71337648
Device Lot Number19BM05339
Was Device Available for Evaluation? No
Date Manufacturer Received06/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-