Catalog Number 71430422
Device Problems Flaked (1246); Material Fragmentation (1261)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/17/2020
Event Type  malfunction  
Event Description
It was reported that during a tka procedure it was found that the genesis ii ps hi flex insert tr s3 4 9 was chipped.The surgery was finished with an s+n backup device.No injuries or surgical delays reported.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A complaint history review found related failures for the listed batch; this failure mode will be monitored for future complaints for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.B5, d4, g1.
 
Event Description
It was reported that during a tka procedure it was found that the genesis ii ps hi flex insert tr s7-8 9 was chipped.The surgery was finished with an s+n backup device.No injuries or surgical delays reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS HI FLEX ISRT TR S7-8 9
Type of Device
TRAY, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10262232
MDR Text Key198482860
Report Number1020279-2020-03129
Device Sequence Number1
Product Code FSM
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 07/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71430422
Device Lot Number07DM01023
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/17/2020
Initial Date FDA Received07/12/2020
Supplement Dates Manufacturer Received07/13/2021
Supplement Dates FDA Received07/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-