• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 028,STERILE,SHELFPAK,; NEBULIZER (DIRECT PATIENT INTE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 028,STERILE,SHELFPAK,; NEBULIZER (DIRECT PATIENT INTE Back to Search Results
Catalog Number 031-28J
Device Problem Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 06/24/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the user found the connection part between the nut adaptor and the flowmeter was unstable before use.Therefore, a new unit was used instead.No patient involvement reported.
 
Manufacturer Narrative
Qn# (b)(4).The sample was returned for evaluation.A visual exam was performed and no defects were observed.Oxygen entrainment testing was performed and during the setup it was observed that the assembly of the nut adaptor and the upper body was unstable.Even with that condition, the sample was able to be tested on oxygen entrainment testing but the testing failed due to the unstable condition.After the testing finished, the adaptor was carefully disassembled from the upper body and it was visually inspected.During the visual inspection, wear was found on the internal tabs.A device history record review was performed and no relevant findings were identified.Based on the investigation performed, the complaint is confirmed.Although the complaint is confirmed, there is not sufficient evidence to assure that this issue was originated during the manufacturing assembly or molding process.The wear on the internal tabs of the adaptor was most likely caused by the end user during the connection of the adaptor into the flowmeter that causes an unstable connection issue.
 
Event Description
It was reported that the user found the connection part between the nut adaptor and the flowmeter was unstable before use.Therefore, a new unit was used instead.No patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON NEBULIZER ADAPTOR 028,STERILE,SHELFPAK,
Type of Device
NEBULIZER (DIRECT PATIENT INTE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key10266793
MDR Text Key199127981
Report Number3004365956-2020-00134
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
PMA/PMN Number
K153010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number031-28J
Device Lot Number74K1900596
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2020
Initial Date Manufacturer Received 06/25/2020
Initial Date FDA Received07/13/2020
Supplement Dates Manufacturer Received07/23/2020
Supplement Dates FDA Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-